top of page

Optical Extinction Analyzers (OEAs) - Specifications*

  • Datasheet

  • Wavelengths: 

    • Standard two different single-wavelength systems available: 520 nm & 1065 nm.

    • Custom single-wavelength system upon customer request: 450 nm, 632 nm, 780 nm, 850 nm, or 1650 nm

  • Precision (see graphs and other information below): 

    • Better than +/- 0.1 Mm-1 (1-sigma, 1s)

    • Better than +/- 0.05 Mm-1 (1-sigma, 10s)

  • Dynamic Range: 0 to 10,000 Mm-1 

  • Data Rate: 1 second (1 Hz) or 0.1 Second (10 Hz) for Eddy Covariance applications - See more here

  • Sample flow rate: 160 SLPM

  • Measurement Technology: Open-Path Cavity Ringdown Spectroscopy 

  • Calibration: Self-Calibrating - No user calibration needed 

  • Consumables: None other than inlet filters 

  • Operating Temperature Range: -5°C to +50°C 

  • Humidity Range: 0 - 99% non-condensing 

  • Size: 49 cm x 33 cm x 18 cm (19.3” x 13” x 7”)

  • Weight: 7.7 kg (17.0 lbs)

  • Power: 35 W - Battery (User-Provided) or wall plug operable

  • Onboard Memory: > 1M readings (2 yrs of continuous data)

  • Output Connectors: USB, HDMI, RS-232

*Subject to change without prior notice

bottom of page